Genotoxicity - Bacterial Reverse Mutation Test with medical devices, cosmetics and disinfectants (ISO 10993-3: 2026 and OECD 471: 2020).

Test Accredited by ENAC (Spanish National Accreditation Entity).

Test with the Certificate of Good Laboratory Practices (GLPs).

The ISO 10993-3standard indicates strategies and tests to identify the risks of genotoxicity, carcinogenicity and reproductive toxicity applicable to assess the potential of medical devices to cause genotoxicity, carcinogenicity and reproductive toxicity. These tests can be performed using mammalian cells, bacteria, yeast or filamentous fungi to determine if the samples tested can induce gene mutations, alter the chromosomal structure or cause other changes in the DNA.

The ISO 10993-1: standard indicates when these tests should be considered in the biological safety evaluation of a medical devices. This standard states that genotoxicity tests are not necessary for medical devices and their components, when it is known that they do not have genotoxicity. The tests are indicated when the materials included in medical devices can have components that can interact with the genetic material, or when the chemical composition of the medical device is unknown.

In accordance with ISO 10993-3, when genotoxicity tests are performed, the test battery must include: a test to detect genetic mutations in bacteria (OECD 471), and an in vitro test with mammalian cells, modified for use with medical devices.

The primary function of genotoxicity testing is to investigate, using cells or organisms, the potential of tested products to induce genetic changes in humans that could be passed on to future generations. Scientific data generally support the hypothesis that DNA damage in somatic cells is critical for the initiation of cancer, and therefore these tests can identify chemicals with carcinogenic potential.

The test described in the OECD 471 (Bacterial Reverse Mutation Test), uses strains of Salmonella typhimurium and Escherichia coli that require the supply of essential aminoacids for their growth, because these strains are unable to synthesize them by being mutants deficient in the metabolic pathways that must synthesize them. When there is a mutagenic compound (genotoxic), this compound can induce the reversion of the mutations, restoring the metabolic functional capacity of the bacteria to synthesize the essential amino acids. Reverted bacteria are detected by their ability to grow in the absence of the aminoacids previously required by the parental test strain.

The performance of this test involves using suspensions of bacteria that are exposed to the substances under evaluation, in the presence or absence of an exogenous metabolic activation system. There are basically two methods of realization: a) plate incorporation method, in which the suspensions are mixed with coating agar and immediately sown in minimal medium; b) pre-incubation method in which the mixture of the bacterial suspension and the substance under test is incubated and then mixed with coating agar to be sown on a minimal medium. In both methods, after 2 to 3 days of incubation colonies should be counted, to detect reverted and compared with colonies reverted spontaneously into control plates. The suspensions of bacteria (5 strains: four of Salmonella spp. and one of Escherichia coli) must be used in the phase of initial exponential growth or initial stationary. The substance under test must be used in at least five concentrations and the corresponding solvent/vehicle controls of the substance under test must be included. Likewise, substances with and without mutagenic capacity must be tested in parallel. The most accepted criterion of mutagenesis is the increase in the detection of mutants depending on the concentration within the tested range and/or reproducible increase at one or more concentrations in the number of reverted colonies per plaque in at least one strain, with or without activation metabolic.

Test phases

The test consists of two phases:

Phase 1 - Previous toxicity test with a bacterium (Salmonella spp. CECT-880) to verify that the product (or that extracted from a medical device) is not toxic or shows interference precipitate at the concentration indicated by the client. If the product (or its extracted is toxic, the test cannot be completed and only this part of the test will be billed).

Phase 2- Bacterial mutation reversal assay with five strains. According to the OECD guidelines, five concentrations of the product must be tested. For solid, non-soluble medical devices, the product extract must be tested. Two extracts must be tested, one in a polar solvent and the other in a non-polar solvent. For medical devices or chemicals, whether liquid or soluble solid, the maximum possible assay concentration should be used. If the product is not cytotoxic and does not form a precipitate, according to ISO 10993-3, testing with the product only at a high concentration or with the pure extract is acceptable. If the product exhibits toxicity, the complete assay with five concentrations must be performed. See the total cost of the test, which varies depending on the type of product being tested.

Test concentration

The recommended maximum concentration according to the standard for soluble non-cytotoxic substances is 5 mg/plate or 5 µl/plate. For non-cytotoxic liquids or extracts, according to ISO 10993-3, the maximum concentration is 100 µL/plate. For cytotoxic products or extracts, a complete assay with 5 concentrations must be performed. Test products that are cytotoxic already below 5 mg/plate or 5 µl/plate should be tested up to a cytotoxic concentration. Testing above the concentration of 5 mg/plate or 5 µl/plate may be considered when evaluating products containing substantial amounts of potentially mutagenic impurities.

The client must indicate if he wishes to perform the test at the maximum concentration indicated by the standard or at another concentration and if the test must be performed at one concentration or 5 concentrations. From the maximum concentration requested by the client, the 4 lower test concentrations to be carried out will be determined, following the interval indicated by the OECD 471 Guideline. Note that if the test product is a non-soluble solid medical device, the test will be carried out only with the pure extract of the product.

Extraction solvents compatible with the test system and metabolic activation must be used; ISO 10993-3 recommends physiological saline solution as polar and DMSO as non-polar.